Articles |
Diagnostic Laboratory, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Ft. Collins 80523, USA.
Twenty forage samples were collected and selected for variation in nitrate content. Each forage samples was analyzed 4 times by 4 different methods: diphenylamine spot plate, spectrophotometric, nitrate-selective electrode, and high-performance liquid chromatographic. Five feed extracts were spiked with 2 different amounts of nitrate and analyzed by each method. The spectrophotometric and nitrate-selective electrode had similar percent recoveries, which were close to 100%. The nitrate-selective electrode method had the least variation of the 4 methods. The diphenylamine spot plate method had the poorest average recovery, greatest variation, and was the least accurate. The average coefficients of variation for all samples within a method were 15%, 12%, 6.4%, and 16 for the diphenylamine spot plate, spectrophotometric, nitrate-selective electrode, and high-performance liquid chromatographic methods, respectively. The variation in the nitrate-selective electrode method was lower (P < 0.05) than the other methods. The results from this study suggest that the nitrate-selective electrode method is more accurate and precise than the other methods of analysis tested.